The “in vitro” study of white thrombus growth by means of Electrical Impedance Spectroscopy on whole blood is presented; the aim of the work is to correlate thrombus formation in an artificial channel with impedance measurements. The measurement bench is composed of a sensor, a switch unit and an impedance meter. Post process of measured data is performed and an equivalent lumped circuit is derived; experimental data are obtained with a first prototype based on a single couple of electrodes.
Electrical Impedance Spectroscopy on flowing blood to predict white thrombus formation in artificial microchannels
AFFANNI, Antonio;SPECOGNA, Ruben;TREVISAN, Francesco
2012-01-01
Abstract
The “in vitro” study of white thrombus growth by means of Electrical Impedance Spectroscopy on whole blood is presented; the aim of the work is to correlate thrombus formation in an artificial channel with impedance measurements. The measurement bench is composed of a sensor, a switch unit and an impedance meter. Post process of measured data is performed and an equivalent lumped circuit is derived; experimental data are obtained with a first prototype based on a single couple of electrodes.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.