The CARdiorenal PEDIatric EMergency (CARPEDIEM) machine was originally designed to perform only continuous venovenous hemofiltration (CVVH) in neonatal and pediatric patients. In some cases, adequate convective clearance may not be reached because of a limited blood flow. In such conditions, the application of diffusive clearance [continuous venovenous hemodialysis (CVVHD)] would help optimize blood purification. In this study, the CARPEDIEM™ machine was modified to enable the circulation of dialysis through the filter allowing testing of the performance of CARPEDIEM™ machine in CVVHD.

CVVHD treatment with CARPEDIEM: small solute clearance at different blood and dialysate flows with three different surface area filter configurations

Vidal, Enrico;
2016-01-01

Abstract

The CARdiorenal PEDIatric EMergency (CARPEDIEM) machine was originally designed to perform only continuous venovenous hemofiltration (CVVH) in neonatal and pediatric patients. In some cases, adequate convective clearance may not be reached because of a limited blood flow. In such conditions, the application of diffusive clearance [continuous venovenous hemodialysis (CVVHD)] would help optimize blood purification. In this study, the CARPEDIEM™ machine was modified to enable the circulation of dialysis through the filter allowing testing of the performance of CARPEDIEM™ machine in CVVHD.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11390/1177182
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 31
  • ???jsp.display-item.citation.isi??? 25
social impact