(1) Background: Infections are among the most frequent and life-threatening complications of cardiovascular implantable electronic device (CIED) implantation. The aim of this study is to compare the outcome and safety of a single-procedure device extraction and contralateral implantation versus the standard-of-care (SoC) two-stage replacement for infected CIEDs. (2) Methods: We retrospectively included 66 patients with CIED infections who were treated at two Italian hospitals. Of the 66 patients enrolled in the study, 27 underwent a single procedure, whereas 39 received SoC treatment. All patients were followed up for 12 months after the procedure. (3) Results: Considering those lost to follow-up, there were no differences in the mortality rates between the two cohorts, with survival rates of 81.5% in the single-procedure group and 84.6% in the SoC group (p = 0.075). (4) Conclusions: Single-procedure reimplantation associated with an active antibiofilm therapy may be a feasible and effective therapeutic option in CIED-dependent and frail patients. Further studies are warranted to define the best treatment regimen and strategies to select patients suitable for the single-procedure reimplantation.

Safety and Efficacy of a Single Procedure of Extraction and Reimplantation of Infected Cardiovascular Implantable Electronic Device (CIED) in Comparison with Deferral Timing: An Observational Retrospective Multicentric Study

Tascini C.;
2023-01-01

Abstract

(1) Background: Infections are among the most frequent and life-threatening complications of cardiovascular implantable electronic device (CIED) implantation. The aim of this study is to compare the outcome and safety of a single-procedure device extraction and contralateral implantation versus the standard-of-care (SoC) two-stage replacement for infected CIEDs. (2) Methods: We retrospectively included 66 patients with CIED infections who were treated at two Italian hospitals. Of the 66 patients enrolled in the study, 27 underwent a single procedure, whereas 39 received SoC treatment. All patients were followed up for 12 months after the procedure. (3) Results: Considering those lost to follow-up, there were no differences in the mortality rates between the two cohorts, with survival rates of 81.5% in the single-procedure group and 84.6% in the SoC group (p = 0.075). (4) Conclusions: Single-procedure reimplantation associated with an active antibiofilm therapy may be a feasible and effective therapeutic option in CIED-dependent and frail patients. Further studies are warranted to define the best treatment regimen and strategies to select patients suitable for the single-procedure reimplantation.
File in questo prodotto:
File Dimensione Formato  
antibiotics-12-01001 (1).pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 756.03 kB
Formato Adobe PDF
756.03 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11390/1255784
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact