Comprehensive two-dimensional gas chromatography (GC×GC) is a powerful technique which can enable a great increase in GC peak capacities. However, since second- ary-column separations are very rapid, detectors with a fast acquisition rate are mandatory. Such a requirement has certainly limited the use of the quadrupole mass spectrometer in the GC×GC field. The present research is focused on the evaluation of a novel rapid-scanning quadrupole mass spectrometry (qMS) detector, charac- terized by a 20 000 amu/s scan speed and a 50 Hz scan frequency, using a 290 amu mass range (40-330 m/z). The performance of the MS system was assessed by analyzing mixtures of 24 allergens, as well as a perfume sample, through GC×GC/qMS. The MS parameters evalu- ated at different acquisition rates (50, 33, and 25 Hz), as well as in the (simultaneous) scan/selected ion moni- toring (SIM) mode, were the number of data points per peak, mass spectrum quality, peak skewing, and sensitiv- ity. Two GC×GC/qMS methods, using the 50 Hz acquisi- tion rate and the scan/SIM mode, were validated. Both methods provided similar results in terms of repeatability, accuracy, and linearity, while a great increase in sensitivity was observed (ca. a factor of 10) under scan/SIM condi- tions. The validated method proved to be suitable for the analysis of perfume allergens, according to the require- ments of Directive 2003/15/EC.

Evaluation of a rapid-scanning quadrupole mass spectrometer in an apolar x ionic-liquid comprehensive two dimensional gas chromatography system

PURCARO, Giorgia;CONTE, Lanfranco;
2010-01-01

Abstract

Comprehensive two-dimensional gas chromatography (GC×GC) is a powerful technique which can enable a great increase in GC peak capacities. However, since second- ary-column separations are very rapid, detectors with a fast acquisition rate are mandatory. Such a requirement has certainly limited the use of the quadrupole mass spectrometer in the GC×GC field. The present research is focused on the evaluation of a novel rapid-scanning quadrupole mass spectrometry (qMS) detector, charac- terized by a 20 000 amu/s scan speed and a 50 Hz scan frequency, using a 290 amu mass range (40-330 m/z). The performance of the MS system was assessed by analyzing mixtures of 24 allergens, as well as a perfume sample, through GC×GC/qMS. The MS parameters evalu- ated at different acquisition rates (50, 33, and 25 Hz), as well as in the (simultaneous) scan/selected ion moni- toring (SIM) mode, were the number of data points per peak, mass spectrum quality, peak skewing, and sensitiv- ity. Two GC×GC/qMS methods, using the 50 Hz acquisi- tion rate and the scan/SIM mode, were validated. Both methods provided similar results in terms of repeatability, accuracy, and linearity, while a great increase in sensitivity was observed (ca. a factor of 10) under scan/SIM condi- tions. The validated method proved to be suitable for the analysis of perfume allergens, according to the require- ments of Directive 2003/15/EC.
File in questo prodotto:
File Dimensione Formato  
ac101678r.pdf

non disponibili

Tipologia: Altro materiale allegato
Licenza: Non pubblico
Dimensione 1.97 MB
Formato Adobe PDF
1.97 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11390/864151
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 86
  • ???jsp.display-item.citation.isi??? 72
social impact